Smart Ventilator and Respiratory Monitoring UI

A Qt/QML-based ICU ventilator user interface demonstration application built with Qt 6.8. The project simulates a real-time respiratory monitoring system with waveform visualization, patient management, alarm handling, and clinical parameter controls. It is designed as a reference implementation for medical device UI development, following separation of concerns between C++ backend controllers and a declarative QML frontend.

Developed by Alsons Technology.

This repository is a ventilator HMI simulator and architecture prototype. It is not cleared or validated for clinical use.

Table of Contents

Overview

This application provides a fully navigable ventilator UI prototype with:

  • Real-time waveform rendering for Pressure, Flow, Volume, and CO2 channels.
  • Simulated patient vitals including SpO2, EtCO2, compliance, and resistance.
  • Eight ventilation mode selections (ASV, SIMV, PCV, CPAP, BiPAP, PSV, PRVC, and related modes).
  • Configurable patient profiles with calculated Ideal Body Weight, tidal volume, and respiratory rate.
  • Active alarm management with critical, warning, and informational severity levels.
  • Persistent settings through QSettings and SQLite database logging.
  • Modular QML component library with centralized theming (colors, typography, spacing, radii).

The simulation engine in VentilatorController generates physiologically plausible waveforms using sine-based modelling. In a production deployment, this simulation layer would be replaced with a hardware adapter communicating over serial, CAN, or Ethernet interfaces while keeping the QML-facing API contract unchanged.

Technology Stack

Layer Technology
Language C++17, QML
Framework Qt 6.8 (QtQuick, QtQuickControls, QtSql)
Database SQLite (via Qt SQL module)
Build System qmake (.pro file)
Documentation Doxygen
Style QtQuick Controls Basic style
Target Desktop (1920x1080 primary, 1366x768 minimum)

Project Structure

MedicalProject/
|-- main.cpp                            Application entry point
|-- main.qml                            Root ApplicationWindow and screen router
|-- MedicalProject.pro                  qmake project configuration
|-- qml.qrc                            QML and asset resource manifest
|-- Doxyfile                            Doxygen configuration
|
|-- src/
|   |-- core/
|   |   |-- AppSettings.h/cpp          QSettings-based persistent preferences
|   |   |-- DatabaseManager.h/cpp      SQLite schema, read/write operations
|   |
|   |-- controllers/
|       |-- AlarmController.h/cpp      Alarm list model and banner state
|       |-- ClockController.h/cpp      Real-time clock (IST timezone)
|       |-- PatientController.h/cpp    Patient demographics and calculations
|       |-- VentilatorController.h/cpp Simulation engine, waveform buffers, alarm evaluation
|
|-- qml/
|   |-- styles/
|   |   |-- Colors.qml                 Color palette singleton
|   |   |-- Typography.qml             Font family and size scale singleton
|   |   |-- Spacing.qml                Layout spacing and margin singleton
|   |   |-- Radius.qml                 Border radius singleton
|   |   |-- qmldir                     QML module registration
|   |
|   |-- components/
|   |   |-- buttons/
|   |   |   |-- PrimaryButton.qml      Styled action button
|   |   |   |-- PrefsTabButton.qml     Tab toggle button
|   |   |
|   |   |-- indicators/
|   |   |   |-- AppHeader.qml          Top bar (mode, patient, clock, alarms)
|   |   |   |-- AlarmBanner.qml        Critical/warning notification banner
|   |   |   |-- DateTimeBanner.qml     Clock and status icon display
|   |   |
|   |   |-- cards/
|   |   |   |-- Panel.qml              Generic styled container
|   |   |   |-- MetricTile.qml         Numeric value display tile
|   |   |   |-- ModeCard.qml           Ventilation mode selection card
|   |   |   |-- StatusPanel.qml        Mode and patient category display
|   |   |
|   |   |-- charts/
|   |   |   |-- WaveformChart.qml      Canvas-based real-time waveform
|   |   |   |-- TrendChart.qml         Trend line wrapper
|   |   |   |-- CircularKnob.qml       Rotary parameter knob
|   |   |   |-- PressureControl.qml    Circular pressure gauge
|   |   |   |-- PressureGroupBox.qml   Gauge with increment/decrement controls
|   |   |   |-- CurvedSideButton.qml   Curved side button for gauge controls
|   |   |
|   |   |-- navigation/
|   |       |-- BottomNavigation.qml   Eight-tab bottom navigation bar
|   |
|   |-- screens/
|   |   |-- SplashScreen.qml           Boot animation with progress indicator
|   |   |-- StandbyScreen.qml          Patient selection and calibration
|   |   |-- PatientSetupScreen.qml     Age, height, weight configuration
|   |   |-- ModeSelectionScreen.qml    Ventilation mode grid
|   |   |-- MonitoringScreen.qml       Live waveforms and vital metrics
|   |   |-- ControlsScreen.qml         Parameter controls (5 sections)
|   |   |-- AlarmCenterScreen.qml      Alarm table with history
|   |   |-- EventsScreen.qml           Event timeline
|   |   |-- SystemDiagnosticsScreen.qml System info, tests, sensors, settings
|   |   |-- ToolsScreen.qml            Utility gauges and alarm log
|   |   |-- LayoutScreen.qml           Layout preset selection
|   |
|   |-- assets/
|       |-- icons/                     SVG and PNG icons
|
|-- docs/
    |-- SCREEN_FLOW_TODO.md            Screen flow documentation and feature roadmap
    |-- PROJECT_STATUS.md              Detailed project status and issue tracker

Architecture

The application follows a controller-view architecture:

flowchart LR
    subgraph QML["QML HMI Layer"]
        Root["main.qml\nLoader screen router"]
        Header["Persistent header\nAlarm + status banners"]
        Screens["Screens\nMonitoring, Controls, Patient, System"]
        Components["Reusable controls\nCards, knobs, charts, navigation"]
    end

    subgraph Cpp["C++ Backend Layer"]
        Vent["VentilatorController\nsimulation + validated command API"]
        Patient["PatientController\nprofile + IBW calculations"]
        Alarm["AlarmController\npriority + history model"]
        User["UserController\noperator sessions"]
        Clock["ClockController\nIndia real-time clock"]
    end

    subgraph Core["Persistence + Platform"]
        Settings["AppSettings\nQSettings"]
        Db["DatabaseManager\nSQLite async writer + audit hash chain"]
        Future["Future hardware adapter\nserial/CAN/Ethernet"]
    end

    Root --> Header
    Root --> Screens
    Screens --> Components
    Screens <--> Vent
    Screens <--> Patient
    Screens <--> Alarm
    Screens <--> User
    Header <--> Clock
    Vent --> Db
    Patient --> Db
    Alarm --> Db
    User --> Db
    Settings --> Root
    Future --> Vent
Loading

C++ controllers are registered as QML context properties in main.cpp. Each controller is a QObject subclass exposing Q_PROPERTY bindings and Q_INVOKABLE methods. QML components bind directly to controller properties for reactive updates.

Screen routing is handled by an asynchronous Loader in main.qml. The currentScreen property selects which Component to load. The bottom navigation bar, clinical alarm banner, and non-clinical system status banner remain persistent across all operating screens.

Style system uses four QML singletons (Colors, Typography, Spacing, Radius) registered via qmldir. All visual properties should reference these singletons rather than using hardcoded values.

Data persistence is split between QSettings (user preferences like brightness, volume, language) and SQLite (alarm history, ventilator snapshots, operator audit events, patient profile changes). DatabaseManager handles schema creation, compatibility migrations, asynchronous writes, and a tamper-evident event hash chain.

Build metadata comes from qmake defines. CI should pass APP_VERSION and BUILD_ID environment variables before running qmake. Local builds default to 0.1.0-dev and local.

Screens

Screen File Description
Splash SplashScreen.qml Boot animation with Alsons Technology branding
Standby StandbyScreen.qml Patient selection, gender, presets, calibration
Patient Setup PatientSetupScreen.qml Configure age, height, weight; shows calculated IBW
Mode Selection ModeSelectionScreen.qml Grid of 8 ventilation modes
Active Monitoring MonitoringScreen.qml Live waveforms, vitals, lung visualization
Controls ControlsScreen.qml Basic, Patient, Advanced, Alarm Limits, Apnea tabs
Alarm Center AlarmCenterScreen.qml Alarm table with severity and acknowledge actions
Events EventsScreen.qml Timeline of mode changes, parameter edits, alarms
System Diagnostics SystemDiagnosticsScreen.qml Device info, self-test, sensors, settings
Tools ToolsScreen.qml Utility gauges and alarm log across 3 pages
Layout LayoutScreen.qml Select monitoring layout presets

Build Instructions

Prerequisites

  • Qt 6.8 or later (with QtQuick, QtQuickControls2, and QtSql modules)
  • C++17 compatible compiler (GCC 9+, Clang 10+, MSVC 2019+)
  • qmake (included with Qt installation)

Build Steps

# Clone the repository
git clone <repository-url>
cd MedicalProject

# Create a build directory
mkdir build && cd build

# Run qmake (adjust path to your Qt installation)
qmake ../MedicalProject.pro

# Build
make -j$(nproc)

# Run
./MedicalProject

To inject build metadata:

APP_VERSION=1.2.0 BUILD_ID="$GIT_COMMIT" qmake ../MedicalProject.pro
make -j$(nproc)

macOS

mkdir build && cd build
/path/to/Qt/6.8.x/macos/bin/qmake ../MedicalProject.pro
make -j$(sysctl -n hw.ncpu)
open MedicalProject.app

Windows (MSVC)

mkdir build && cd build
C:\Qt\6.8.x\msvc2019_64\bin\qmake ..\MedicalProject.pro
nmake
MedicalProject.exe

Configuration

Application settings are persisted through QSettings under the organization "AlsonsTechnology" and application name "SmartVentilatorDemo". Configurable values include:

Setting Type Default Description
softwareVersion QString "1.0.0" Read-only software identifier
operatingHours double 0.0 Accumulated runtime hours
brightness int 80 Display brightness (0-100)
audioVolume int 50 Audio level (0-100)
language QString "en" UI language code

Documentation Generation

The project includes a Doxyfile for generating API documentation from source code comments.

# Generate HTML documentation
doxygen Doxyfile

# Open generated docs
open docs/html/index.html

Embedded Deployment

Deployment starter files are provided under deploy/:

  • deploy/systemd/screenshots-ui.service for supervised embedded launch.
  • deploy/yocto/screenshots-ui.bb as a Qt 6 Yocto recipe template.
  • docs/EMBEDDED_DEPLOYMENT.md for filesystem, watchdog, and runtime safety notes.

Design Reference

The UI design is based on the Smart Ventilator and Respiratory Monitoring UI concept by Tahir Moosani. The full design specification can be viewed at:

https://www.behance.net/gallery/211931131/screenshots-Respiratory-Monitoring-UI

Key design principles from the reference:

  • Dark theme with blue accent palette for reduced eye strain in ICU environments.
  • High-contrast text for readability under variable lighting conditions.
  • Large touch targets for gloved-hand operation.
  • Clear visual hierarchy separating waveforms, vitals, and controls.
  • Color-coded severity levels for alarm states.

Qt Licensing

Qt module and licensing considerations are tracked in docs/QT_LICENSING_NOTES.md. A closed medical embedded product should complete a formal Qt LGPL/commercial licensing review before release.

Coding Standards

QML

  • Use Qt 6 versionless imports (e.g., import QtQuick not import QtQuick 2.15).
  • Reference style singletons (Colors, Typography, Spacing, Radius) for all visual values.
  • Add pragma ComponentBehavior: Bound to files using delegates or Repeaters.
  • Maximum line length: 120 characters.
  • Use 4-space indentation throughout.
  • Order QML properties: id, required properties, custom properties, signals, signal handlers, functions, child items.
  • Document component interfaces with header comments.

C++

  • Use #pragma once for header guards.
  • Follow Qt naming conventions (camelCase for methods, m_ prefix for member variables).
  • Document public API with Doxygen comments (@brief, @param, @return).
  • Use QStringLiteral for compile-time string construction.
  • Validate all external inputs with qBound in property setters.

Medical Device Considerations

This is a demonstration application and is not certified for clinical use. A production medical ventilator UI would require additional compliance work including but not limited to:

  • IEC 62304: Medical device software lifecycle processes.
  • IEC 60601-1-8: Alarm system requirements for priority, escalation, and silence timing.
  • IEC 62366-1: Usability engineering for medical devices.
  • FDA 21 CFR Part 11: Electronic records and audit trail requirements.
  • WCAG 2.1 AA: Accessibility including color-blind safe indicators.
  • Risk management per ISO 14971 with documented hazard analysis.
  • Minimum touch target sizes of 44x44 pixels for gloved operation.
  • Screen lock and operator authentication for clinical environments.
  • Tamper-proof audit trail logging for all parameter changes and alarm events.

See docs/PROJECT_STATUS.md for the current compliance gap analysis and implementation roadmap.

Smart Ventilator Project Screens

The public/screenshots folder contains the complete ICU Smart Ventilator UI screen set. This project is a Qt/QML clinical HMI concept for a bedside ventilator: it covers startup checks, secure operator access, patient setup, live waveform monitoring, alarms, trends, therapy tools, weaning support, clinical records, export, maintenance, and device settings.

The design is built around a simple idea: an ICU operator should always know what mode the device is in, who the patient is, what alarm is active, and where to go next. Critical actions stay visible in the header, while the bottom navigation keeps the main workflows predictable.

Project demo video: Smart ICU Ventilator UI on YouTube

1. Boot Self-Check

Boot Self-Check

The ventilator opens with a branded startup screen for Alsons Technology. It shows sensor-check progress, software version, and total operating hours, which makes the device feel like a real embedded medical system rather than a simple mock screen.

2. Operator Login

Operator Login

The login screen keeps the operator focused on authentication. The username field, PIN indicators, numeric keypad, clear button, and OK button are large enough for a touch interface and avoid unnecessary distractions.

3. Filled Login State

Filled Login State

This screen shows the entered operator name and filled PIN dots. It confirms the login interaction state clearly before the operator submits credentials.

4. Standby And Patient Start

Standby And Patient Start

The standby screen clearly warns that no ventilation is being delivered. From here the operator can select neonatal/adult profiles, choose gender, set patient height, run test and calibration, and start ventilation. Oxygen, PEEP/CPAP, and minute-volume target controls stay available on the right.

5. Patient Profile Configuration

Patient Profile Configuration

This screen turns patient data into suggested settings. Age, height, weight, patient category, and gender update predicted body weight, tidal volume, and respiratory rate so the operator can review a sensible starting point before continuing.

6. Live Monitoring With High-Pressure Alarm

Live Monitoring With High-Pressure Alarm

The main monitoring view combines live pressure, flow, volume, and PCO2 curves with measured values on the left. A high-pressure alarm is visible at the top, while quick adjustments remain available on the right.

7. Driving Pressure Warning

Driving Pressure Warning

This state shows a high driving pressure alert with a practical warning: reduce tidal volume or increase PEEP. The UI does more than flash an alarm; it gives the operator a useful next thought.

8. Basic Ventilation Controls

Basic Ventilation Controls

The basic controls tab groups the most common ventilation parameters: FiO2, PEEP, pressure support, respiratory rate, trigger, tidal volume, and minute-volume target. Circular controls and plus/minus actions make touch adjustment straightforward.

9. Patient Controls

Patient Controls

The patient controls screen lets the operator review ventilation time, patient height, gender, and ideal body weight. The save profile action keeps patient-specific values tied to the current ventilation session.

10. Advanced Controls

Advanced Controls

Advanced settings are separated from the everyday controls. Ramp, oxygen, pressure limit, PEEP/CPAP, ETS, and minute-volume target can be tuned without crowding the monitoring screen.

11. Alarm Limit Controls

Alarm Limit Controls

Alarm thresholds are grouped in one place: high pressure, low pressure, apnea time, low tidal volume, high minute volume, and low SpO2. This makes safety configuration easy to review before and during therapy.

12. Apnea Backup Ventilation

Apnea Backup Ventilation

Apnea backup is treated as its own safety workflow. The operator can turn backup on, see the selected mode, and configure backup rate, backup tidal volume, and backup PEEP.

13. One-Hour Trends

One-Hour Trends

The trends page gives a compact one-hour history for Ppeak, SpO2, EtCO2, FiO2, PEEP, and static compliance. It is designed for quick bedside review.

14. Six-Hour Trends

Six-Hour Trends

The six-hour view helps operators see how settings and patient response changed over a longer shift window.

15. Twenty-Four-Hour Trends

Twenty-Four-Hour Trends

The 24-hour trend view supports handover and retrospective review. The latest value stays visible for each chart, so long-term context does not hide the current state.

16. Respiratory Loops

Respiratory Loops

Respiratory loops show pressure-volume and flow-volume behavior in real time. The freeze button lets a clinician pause the curves for closer inspection.

17. Clinical Admission Record

Clinical Admission Record

The clinical section starts with patient admission data: patient ID, bed number, physician, and admit date. Admit/update and discharge actions are clearly separated.

18. Therapy Controls

Therapy Controls

Therapy controls include the heated humidifier and nebulizer. The screen shows target and actual humidifier temperature, water level, medication name, and nebulizer duration.

19. Weaning Readiness

Weaning Readiness

The weaning screen brings together RSBI, SpO2, PEEP, FiO2, work of breathing, stress index, dead-space ratio, and oxygen time. It also gives a clear start action for a supervised breathing trial.

20. Respiratory Maneuvers

Respiratory Maneuvers

Inspiratory hold and expiratory hold are presented as clinical maneuvers with a history log. This helps make plateau pressure and auto-PEEP checks traceable.

21. Clinical Data Export

Clinical Data Export

The export screen supports parameter snapshots and audit-event export. This is useful for clinical handover, technical review, and offline reporting.

22. Clinical Reference

Clinical Reference

The reference area keeps clinical guidance close to the operator workflow, reducing the need to leave the ventilator interface for common reference checks.

23. Network Status

Network Status

Network status is separated from ventilation controls. That makes connectivity and communication checks available without adding risk to active therapy settings.

24. Maintenance Overview

Maintenance Overview

Maintenance screens keep service and device-health workflows away from bedside controls, while still keeping the current mode and alarm context visible.

25. Central Monitoring

Central Monitoring

Central monitoring supports a broader ICU workflow where ventilator status may need to be reviewed beyond the local device screen.

26. System Status

System Status

The system section presents device status information such as power, battery, runtime, and network readiness in one place.

27. Battery And Network Detail

Battery And Network Detail

Battery and network indicators are kept visible because they affect whether the ventilator can continue operating reliably.

28. Layout Configuration

Layout Configuration

The layout screen supports different monitoring arrangements while preserving the same global navigation and header structure.

29. Target Settings

Target Settings

Target settings centralize goal-based ventilation values so the operator can tune therapy without jumping through unrelated menus.

30. Alarm History

Alarm History

Alarm history helps staff understand recent events before the current state. This is especially useful during shift changes and post-event review.

31. Tools Overview

Tools Overview

Tools collect setup, checks, and troubleshooting actions in a dedicated area so normal monitoring remains clean.

32. Ventilation Modes

Ventilation Modes

The modes screen lets the operator review and select ventilation strategies inside the same interface system used during active therapy.

33. Settings Overview

Settings Overview

Settings collect device-level configuration and preferences without mixing them into clinical parameter screens.

34. Mode Change Confirmation

Mode Change Confirmation

Mode changes are high-impact actions, so the interface uses confirmation states to reduce accidental changes.

35. Emergency Control State

Emergency Control State

The emergency action stays prominent in the top bar. This screen shows how the interface keeps critical actions reachable during abnormal conditions.

36. Power-Off Confirmation

Power-Off Confirmation

Power-off is handled as a deliberate action instead of a casual button press, which is important in a clinical device.

37. Frozen Waveform Review

Frozen Waveform Review

Freeze mode lets clinicians pause waveform movement and inspect the latest pressure, flow, volume, and CO2 behavior.

38. Resumed Waveform Monitoring

Resumed Waveform Monitoring

The resume state returns the operator to live monitoring while preserving the same values, alarm context, and quick controls.

39. Alarm Muted State

Alarm Muted State

Muted or acknowledged alarm states still keep the warning visible, which helps reduce alarm fatigue without hiding risk.

40. Patient Category Switching

Patient Category Switching

Adult, pediatric, and neonatal patient choices are shown as clear segmented actions so the UI can adapt suggested values to the patient group.

41. Test And Calibration Flow

Test And Calibration Flow

The test and calibration entry point sits close to standby operation, which is where operators need setup checks before ventilation begins.

42. Suggested Settings Review

Suggested Settings Review

Suggested values are shown before the operator continues to modes, making the relationship between patient inputs and starting parameters easy to review.

43. High-Pressure Recovery

High-Pressure Recovery

Recovery states keep alarm language beside live waveform feedback, helping operators judge whether a setting change is improving the situation.

44. Service Workflow

Service Workflow

Service-focused screens are discoverable but still preserve the clinical header, alarm state, and device identity.

45. Complete Operator Console