Rust rewrite v1 skeleton: regulator-core + ratatui TUI - #1
Merged
Merged
Conversation
Port of tg's timing algorithm (noise suppression, FFT autocorrelation period detection, waveform folding, beat error, amplitude) to a pure-Rust core crate with unit tests against a synthetic watch generator, plus a clean three-zone terminal UI with cpal capture (optional feature) and a --demo mode. Original C code kept as reference. Co-authored-by: borosf <borosf@users.noreply.github.com>
…esample to 44.1 kHz mono Co-authored-by: borosf <borosf@users.noreply.github.com>
Co-authored-by: borosf <borosf@users.noreply.github.com>
Co-authored-by: borosf <borosf@users.noreply.github.com>
…aching, egui GUI - TimingResult fields are now Option: rate/amplitude/beat error report independently and blank out when not found - Amplitude: direct impulse-duration measurement (region above 20% of the tic/toc peak) replacing tg's fragile escalating-threshold heuristic; validated at 264 deg vs 270 deg injected on the synthetic watch - Period: parabolic interpolation of the autocorrelation peak for sub-sample resolution - Analyzer caches FFT plans per length and reuses scratch buffers - New regulator-gui crate: egui frontend with live plot, settings panel, demo/mic sources shared with the TUI - TUI waveform renderer: proper scaling and tic/toc marker Co-authored-by: borosf <borosf@users.noreply.github.com>
Real microphone noise can collapse the autocorrelation search window to a==b or run it past the buffer end; tg's C loops handle this implicitly but the Rust port sliced out of range. Guard the ranges, reject right-edge peaks like the C code does, and make the frontends resilient to a poisoned lock. Regression tests for the degenerate cases. Co-authored-by: borosf <borosf@users.noreply.github.com>
- regulator-core::calibrate: CalibrationState ports tg's calibration (quartz watch ticking 1/s, phase drift fit over CAL_DATA_SIZE=900 one-second samples); Analyzer applies the correction to rate (measured - calibration); validated on synthetic quartz: -15.2 s/d recovered for a +15 s/d fast clock - GUI: calibrate button with live progress, calibration readout, calibration runs on mic input with sliding 4 s windows - analyze_wav example: analyze any WAV recording, printing per-block results, for debugging against real recordings Co-authored-by: borosf <borosf@users.noreply.github.com>
…d-trip tests Co-authored-by: borosf <borosf@users.noreply.github.com>
…gorithm Debugged against a real 18000 bph watch recording (previously: no signal, +5000 s/d garbage): - tg's default filter chain high-passes at FILTER_CUTOFF (3000 Hz) before analysis; without it, mains hum dominated the autocorrelation and the period detector failed on most blocks. Ported the biquad high-pass. - Restored tg's exact amplitude algorithm (escalating-threshold scan for the pre-impulse build-up): measures 236-238 deg on the real recording where tg shows 272 deg. The earlier 20%-span variant was mistuned. - Synthetic generator now models the escapement build-up (pre-hump) so the amplitude path works in demo mode; quartz mode keeps sharp clicks for calibration. Results on the real recording: rate +40..+52 s/d (tg: +33), amplitude 237 deg (tg: 272), beat error ~0.1 ms, bph auto-detected 18000. Co-authored-by: borosf <borosf@users.noreply.github.com>
- TimingResult now carries located tick events (fractional sample positions, tic/toc flag), extrapolated from the folded waveform peaks through the block at the detected period - GUI: live Trace panel under the waveform - each tick plotted at its phase error over the last 60 s; rate error drifts as diagonal bands, beat error splits into tic/toc bands, noise is scatter - Calibration refits on every new one-second sample once 60 points are collected and stops as soon as the confidence interval is under 0.1 s/d, instead of always waiting for tg's full 900 s Co-authored-by: borosf <borosf@users.noreply.github.com>
…eadout - Strip chart is now the centerpiece and vertical like a timegrapher: time flows downward (newest at the bottom), horizontal axis is the phase error against the beat grid (+/-25 ms window, wrapping), tic green / toc magenta dashes, grid lines every 10 ms - Layout per instrument UX: one large monospace rate readout at top (color-coded), secondary values (amp/be/lift/noise/cal) in a compact bottom strip, controls in a slim side panel, small waveform in the panel - no more three-equal-cards grid - Beat rate dropdown extended to all standard rates 12000-43200; 36000 bph covered by a new auto-detect test Co-authored-by: borosf <borosf@users.noreply.github.com>
- During calibration the center panel becomes live feedback: collected (time, phase) samples plotted, progress counter, current estimate with its standard error, and a 'waiting for ticks' hint when no quartz signal is detected - Calibrate button now switches from demo to mic automatically instead of silently doing nothing - Lift angle is a typed DragValue (30-70) instead of a slider - 'export chart' button writes the strip chart as SVG plus raw points as CSV (timestamped filenames in the working directory) - Removed the license footer label Co-authored-by: borosf <borosf@users.noreply.github.com>
The 900-sample quartz calibration exists because tg averages phase noise against a wristwatch reference. But the sound card's clock error can be measured directly against the NTP-disciplined system clock: regress the cumulative captured sample count against batch arrival times (regulator-core::ClockCal). This runs passively whenever the mic is on - no watch, no mode, no user action - and converges within a few minutes; the fit's standard error shrinks as batches accumulate and the correction auto-applies once it is under 0.25 s/d (shown as 'cal auto' in the strip). The quartz path remains available as an independent check. Co-authored-by: borosf <borosf@users.noreply.github.com>
Co-authored-by: borosf <borosf@users.noreply.github.com>
…readout - Events are now measured, not extrapolated: each expected tick position is searched in the prepared signal for the actual peak (rejected if it does not rise above the local noise floor), so the trace shows genuine per-tick scatter and drift instead of static stripes - Mic ring buffer carries a monotonic sample counter, so window start times no longer oscillate when the buffer drains (the cause of the frozen, patternless chart); demo mode uses a running total too - Clock calibration shows live while converging: 'cal +2.1 +-1.4' in amber in the bottom strip until it auto-applies as 'cal ... auto' Co-authored-by: borosf <borosf@users.noreply.github.com>
…e draw - Worker copies the 4 s window out of the ring buffer under the lock (sub-millisecond) and analyzes it unlocked, so the capture relay and UI never stall behind analysis - One Analyzer is reused across iterations, keeping the FFT plan cache warm instead of rebuilding 350k-point plans every 200 ms - Export button moved into the top readout bar; the egui panel added after CentralPanel is gone (panels must precede the central panel) - draw_trace hoists the newest-timestamp lookup out of the per-point loop (was O(n^2) per frame) Co-authored-by: borosf <borosf@users.noreply.github.com>
The clock-calibration block locked params a second time while the outer guard was still held; std::sync::Mutex is not reentrant, so the worker hung forever the moment the clock estimate first became available (~2 s after the mic started), taking the UI with it. Reuse the held guard instead. Co-authored-by: borosf <borosf@users.noreply.github.com>
The clock-based estimate measures the sound card against the local high-resolution timer, whose own drift is unknown and can be tens of s/day on some hardware (observed: -31.9 vs tg's quartz -7.5, a ~28 ppm local timer drift). It is no longer auto-applied: the bottom strip shows it as 'clock ...' in gray, and a side-panel button lets the user apply it explicitly. The quartz calibration remains the authoritative correction, matching tg. Co-authored-by: borosf <borosf@users.noreply.github.com>
- Quartz calibration accepts at +/-0.2 s/d confidence (tg gates at 0.1), roughly halving convergence; the 900-sample count is only the worst-case cap, the early-exit fit typically finishes in a minute or two - Calibration is a typed, editable field in the side panel: a value measured once (here or in tg) can be entered directly - Settings (beat rate, lift angle, calibration, noise suppressor) now persist to %APPDATA%/regulator/config.txt and reload at startup, so calibration survives restarts; tg's startup 'auto cal' only catches device-forced rate mismatches and is ~0 in practice, so there was nothing faster to inherit Co-authored-by: borosf <borosf@users.noreply.github.com>
…ME, CONTRIBUTING) - Trace reworked to tg's orientation: time left-to-right, phase error vertical; SVG export matches. - New branded JPEG result sheet (regulator-gui/src/sheet.rs) with a hand-drawn 5x7 bitmap font, readouts, trace and timestamp; exported alongside SVG/CSV. Unit-tested. - snap_bph now includes 19800 and 43200 (auto mode previously snapped a 19800 bph watch to the wrong standard). - Event location calls vmax once instead of twice per tick. - README rewritten (GUI is the main frontend, TUI marked experimental), screenshot from the sheet renderer. - CI (ubuntu+windows full, headless build), release workflow attaching GUI binaries on v* tags, CONTRIBUTING.md, bug report template. Co-authored-by: borosf <borosf@users.noreply.github.com>
This file contains hidden or bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
Summary
regulator-core: pure-Rust port of tg's timing algorithm — impulsive-noise suppression, FFT-based autocorrelation period detection (auto or fixed bph), waveform folding, beat error, and lift-angle amplitude estimation. No I/O, no platform deps.regulator-tui: clean three-zone terminal UI (readout / waveform / status), live capture via cpal behind the optionalaudiofeature, and a--demomode driven by a synthetic watch signal so it runs with no hardware or audio stack.docs/original-readme.md.Verification
cargo test— 4 passedcargo build --no-default-features— clean (sandbox has no ALSA, so the cpal feature is verified by config only)Notes
Noneon the synthetic signal (its pulse shape doesn't fully match a real tic/toc); shows as---in the UI. Needs validation against real recordings.