Fluke 19x ScopeMeter tools

Caution

Use at your own risk. These tools write to the meter's flash memory and calibration data. A failed or interrupted write, a bad contact, a wrong option or a bug in these tools can brick the meter, destroy its factory calibration, or leave it reading incorrectly.

Make a full backup before you change anything, check that it reads back correctly, and keep it somewhere safe. A lost calibration can't be recreated without a professional recalibration.

This software is provided "as is", without warranty of any kind, express or implied. The author accepts no responsibility or liability for bricked or damaged meters, lost calibration, inaccurate measurements, voided warranties, damage to other equipment, or any other direct or indirect loss resulting from the use or misuse of these tools or the information in this README. By using them you accept full responsibility for the outcome.

This project is not affiliated with, endorsed by or supported by Fluke Corporation. Fluke and ScopeMeter are trademarks of Fluke Corporation. Modifying a meter may void its warranty and calibration certification; a converted meter's calibration is not traceable to any 199C calibration.

Python tools for the Fluke 196/199 (B/C) ScopeMeters running firmware V08.04:

  • enable the Medical and BusHealth options over the optical serial port,
  • convert a 196B/196C memory module into a 199C while keeping the meter's own calibration,
  • dump the 2 MB flash over the optical port,
  • grab a trace and plot it with its FFT,
  • build a patched V08.04 image with the sample rate and a Li-ion battery gauge.

Requires Python 3 with pyserial; scope_fft.py also needs numpy and matplotlib.

pip install pyserial numpy matplotlib

The serial-port tools need only the meter's optical cable. The memory-module conversion also needs the fluke19x flash programmer from feedbackloopchannel/fluke19x-programmer: an ATmega3290 board with a card-edge socket for the meter's memory module. Build it and load its firmware as described there; these tools talk to it at 115200 baud.

Back up first. Every tool that writes saves a backup before it changes anything. Keep those .bin files: each one holds a meter's calibration, which can't be recreated without a recalibration.

Tools

Tool Needs What it does
enable_all_options.bat → tools/enable_all_options.py optical cable Turns on Medical + BusHealth
convert_to_199c.bat → tools/convert_to_199c.py programmer + module Converts a 196B/196C module to 199C V08.04, keeping its cal
tools/serial_dump.py optical cable Full 2 MB flash dump
tools/scope_fft.py optical cable Captures input A and plots waveform + FFT
firmware_patches/build_image.py programmer + module Builds a patched V08.04 image: sample rate and Li-ion battery gauge (see Firmware patches)

The .bat files are double-click wrappers for Windows; they pass any arguments through and keep the window open at the end.

enable_all_options

python tools/enable_all_options.py [port] [--yes]
  1. Tries every serial port (or the one given) at 115200, 1200 and the other standard speeds, sending ID until a meter answers.
  2. Refuses anything that isn't a 19xC on V08.04, since the option codes are only known for that version.
  3. Enters maintenance mode, reads both cal banks with RB and saves them as cal_backup_<serial>_<time>.bin.
  4. If both options are already on, stops. Otherwise asks, then sends CI 17,161027 and RC.
  5. Reads the cal back and checks that the options are set and the checksum is good.

RC saves the cal record, and it also sets the cal date to today and adds 1 to the cal counter. No calibration actually takes place.

convert_to_199c

python tools/convert_to_199c.py [--port COMx] [--donor donor_199c_v0804.bin] [--color-lcd] [--options all|keep] [--force-st] [--yes]
python tools/convert_to_199c.py --source readback.bin --out image.bin [--donor ...]   # offline build only
python tools/convert_to_199c.py --restore backup.bin [--port COMx] [--force-st]       # write a backup back

The donor is a full 2 MB dump of any 199C running V08.04. Its bootloader and code are used, never its calibration. By default the tool looks for donor_199c_v0804.bin in the repo root.

  1. Finds the programmer on any port by asking for flash chip IDs. SST (0x00BF/0x2781) chips are written directly; ST M29W800 (0x0020/0x225B) chips need --force-st (see Flash chips).
  2. Reads the module twice. If the reads differ (bad contact), it stops. Otherwise it saves module_backup_<serial>_<model>_<time>.bin.
  3. Builds the image from the donor's bootloader and code plus the module's own cal banks, then patches both banks: model string 199C, the 196→199 flag at 0xDA, and (with --options all, the default) both options. A mono 196B also gets the color-display values, but only if you pass --color-lcd to confirm a color LCD has been fitted. If any value isn't one it expects, it refuses to convert.
  4. Prints exactly which cal offsets changed and asks before going on.
  5. Erases the module, writes it (3–4 minutes) and verifies with a full readback.

--restore writes any 2 MB file back unchanged, with the same erase/write/verify steps. Run long writes with python -u so the progress output isn't buffered.

serial_dump

python tools/serial_dump.py <port> <out.bin> [baud]

Enters maintenance mode, reads 0x40000000–0x401FFFFF with RB in 16 KB blocks (retrying bad blocks), then sends EO. Nothing is saved to the meter. At 115200 baud it takes about 7 minutes. The meter has to be at that speed already (see Optical serial port).

scope_fft

python tools/scope_fft.py <port> [--baud 115200] [--trace 10] [--avg 8] [--window hann] [--csv out.csv] [--png out.png] [--no-show]

Captures a trace with QW (10 = input A), converts the samples to volts using the scale data the meter sends, and plots the waveform and a power-averaged FFT.

After a 196B → 199C conversion

  • Boot values. A converted 196B only boots with the 199C bootloader values at 0x74–0x7F (see Flash layout). convert_to_199c.py takes them from the donor.
  • Don't update it with Fluke's FlashTool. The official serial update rewrites 0x74–0x7C with 196-type values, so even a successful update leaves a converted 196B unbootable. See Firmware updates over the optical port.
  • Display contrast. A color LCD fitted to a 196B starts with the mono LCD's contrast settings. Set them in the meter's cal mode, steps CL 0100 (max dark), CL 0110 (default) and CL 0120 (max bright), as in the service manual, section 5.4. To get there: hold USER, press and release CLEAR MENU, release USER. F1 steps to CL 0100; F3 starts and ends each step, the arrows adjust, F2 moves to the next step; F4 EXIT, YES saves. Per section 5.7, changing only the contrast doesn't change the cal date or counter. Do it on the charger, and don't press F3 at the WarmingUp step (CL 0200).
  • Bandwidth. Above 100 MHz, a converted 196 keeps its original 100 MHz calibration. Get a 199C calibration if accurate high-frequency readings matter.

Firmware patches

firmware_patches/ builds a patched image from your own full 2 MB dump and adds three things:

  • the sample rate on screen,
  • the battery voltage next to the battery symbol,
  • a battery gauge that suits a 2S Li-ion pack.

No firmware is included. Only the scripts are here, and their output (*.bin) is kept out of the repo.

Warning

  • V08.04 only. The patches hook fixed addresses in the V08.04 code. Before changing anything, build_image.py checks the original word at every patch site and checks that the code area is erased. If anything differs, it stops.
  • Write it with the programmer only (convert_to_199c.py --restore). Fluke's FlashTool file doesn't include the patch code, and FlashTool would make a converted 196B unbootable (see Firmware updates).
  • The battery part is tuned for one pack and one meter (below). Don't use it with a stock NiMH pack.
  • Tested on one meter (ex-196B converted to 199C, 2S2P NCR18650 pack). It runs there as v7; v8 (the * rate marker) is tested in the emulator only.
pip install capstone unicorn
cd firmware_patches
python build_image.py <your_dump.bin> <patched.bin>      # writes patched.bin and patched.bin.txt (patch list + listings)
python emutest.py <your_dump.bin>                        # battery code in an ARM emulator: should end "0 failures"
python toptest.py <your_dump.bin>                        # top-band code: should end "0 failures"
python -u ../tools/convert_to_199c.py --port COMx --restore <patched.bin>   # add --force-st for an ST module

build_image.py never overwrites an existing output. Use a dump of the module you're about to write: the cal banks are copied unchanged from it. The optional third argument picks an older version (v7, v6, top = v5, top3 = v3, 84 = v2, 279 = v1). Those rebuild byte for byte and exist for reference; v8 is the default.

What each patch does

New code goes into erased flash at 0x1E9000–0x1E9D73. The firmware reaches it through repointed entries in the UI callback table (0x401AF29C, 360 entries) and a few redirected branches.

Patch File What it does
Sample rate + voltage in the top band toppatch.py, samplerate.py Callback #181 (status text ending in the battery glyph) shows 2.5G 8.23V<glyph> on the plain scope screen. When the rate has 3+ digits it shows one voltage decimal (1.25G 8.2V), so the text fits the ~5-glyph box. The rate is 1/(time per trace point), 25 points per division. Above the 2.5 GS/s real-time ADC rate (10 ns/div), the trace points are equivalent-time or reconstructed, so v8 adds a * (5G* at 5 ns/div). On every other screen the text stays stock: meter mode, spectrum, bus health, Scope Record and any status text. Callback #84 (bottom-bar timebase) stays stock, but it triggers a redraw so the rate follows timebase changes.
Battery voltage battpatch.py (cb181) Puts the pack voltage, X.YZV, in front of the battery/plug glyph.
Coulomb-counting gauge battpatch.py (integ) Replaces the firmware integrator. The stock one counts charge from the current setpoint, which overcounts badly once a voltage clamp limits the current. The new one counts the measured current both ways (99 % charge efficiency) and remembers any shortfall below empty. It marks "full" on the constant-voltage taper and redraws the voltage text when it moves by 10 mV.
Capacity learning battpatch.py (learn) At the empty event, if a full mark was seen since the last one, capacity = charge drawn between full and empty, clamped to 1000–11000 mAh. This replaces the stock +100 mAh step.
Gauge start from voltage battpatch.py (estimate, binit) After a battery reconnect, the firmware resets the gauge to capacity/4. The patch replaces that with an estimate from the resting voltage, with load compensation.
Firmware tweaks battpatch.py (build) No 6 h-timer or dT/dt "full" marks. No −2 mAh on charger connect. Self-discharge 1 %/day → ~0.1 %/day. Low-battery warning 6.9 V → 6.6 V. Initial capacity 6000 mAh.

The integrator's variables live in high RAM at 0x440FFF00: magic BAT2, deficit, anchored flag, taper counter, last mV, capacity copy, estimate pending. They reset when the battery is disconnected.

Pack-specific settings

All in the marked block at the top of battpatch.py, plus the thresholds listed there:

Setting Value here Why
INIT_CAP 6000 mAh 2S2P NCR18650. Learning corrects it after one full→empty cycle
MIN_CAP, MAX_CAP 1000, 11000 mAh Keep MAX_CAP ≤ 11000: the firmware's 95 % compare overflows above ~12500
SOC_CURVE 8250 mV = 100 % … 7580 = 50 % … 6000 = 0 2S Li-ion resting voltage, with full matched to the ~8.26 V hardware clamp
V_CORR +80 mV This meter's ADC reads 80 mV under a DMM. Measure yours
SOC_RLOAD 0.2 Ω Load compensation for the voltage estimate
Full detection meter off, 0 ≤ I < 100 mA, 2·I < setpoint, V > 7.9 V for ~5 min Constant-voltage taper; doesn't depend on the exact clamp voltage
Low warning 6.6 V Stock 6.9 V is too early for Li-ion. Empty stays at 6.0 V

The patch does not limit the charge voltage. The firmware still charges with its NiMH profile, so the pack needs a hardware clamp (the R4113 change in Battery and charger) and a BMS. Temperature limits are stock.

How it works

Optical serial port

8N1, every command ends in CR. Each command gets a one-character reply: 0 OK, 1 syntax error or not allowed right now, 2 bad value.

The meter starts at 1200 baud after every off/on. Send PC 115200 at 1200 and work at 115200: at 1200 the replies to back-to-back commands overlap and maintenance mode can be refused. PC 1200 sets it back. If a script dies mid-transfer, wait a few seconds and send a bare CR before retrying.

Command Use
ID Model; firmware version
PC <baud> Change the port speed
EM, then maintenance Enter maintenance (calibration) mode
EO Leave maintenance mode. Nothing is saved without RC
RB <addr>,<count> Read memory (decimal arguments; flash starts at 1073741824 = 0x40000000). Reply is #H followed by the hex
CI 17,<code> Option code: 271267 Medical, 110195 BusHealth, 161027 both, 0 none
RC Record calibration: writes the cal record, sets cal date to today, adds 1 to the cal counter
QW <trace> Waveform data
QP Screen print: Epson ESC * mode-4 stripes, 352×240, 1 bit per pixel (colors are thresholded)

Never send EM with flukeuhm unless you mean to flash the meter: it switches to the bootloader's update mode (see below).

The other commands the firmware accepts: AS AT CL CM CV DS EX FM GD GL GR IS PS QC QI QM QP QS RD RI RS RT RW SO SS ST TA VE WB WC WD WT WW RN HO CD DT RP IM CN PW CC KY. Each one is only allowed in certain meter states; the per-command masks are at ROM 0x401A4794.

The front-panel BusHealth install code for V08.04 is 3 × serial + 161072.

If the port hangs, reseat the memory module first: a slightly loose module caused repeated hangs on plain RB reads. Long bulk reads of RAM (around 0x4402D000) also hung the meter, which recovered only after both the charger and the battery were removed. Keep RAM reads small.

QW reply

0\r, then #0 0x80 u16 length, the admin block and its checksum, then ,#0 0x90 u32 length, the samples block and its checksum. Each checksum is the 8-bit sum of its payload. Admin values are 3-byte numbers (i16 mantissa, i8 power of ten): V/div, s/div, y zero, x zero, V per count, s per sample. The samples block holds a format byte (0x40 = min/max pairs, low nibble = bytes per sample), overload/underload/invalid codes, a u16 count, then the samples. Volts = count × V-per-count + y zero.

Flash layout

2 MB, as 32-bit words across two x16 chips, mapped at 0x40000000.

Range Contents
0x000000–0x007FFF Bootloader (boot header, magic 'TEIN', at 0x00–0x27)
0x008000–0x00BFFF Cal/config bank A
0x00C000–0x00FFFF Cal/config bank B
0x010000–0x1E8667 Firmware code (V08.04)
0x1E8668–0x1EFFFF Erased in the stock image
0x1F0000–0x1FFFFF Reserved; leave alone

The V08.04 code is identical for the 196C and 199C. The model and its options come from the cal block, plus the bootloader bytes at 0x74–0x7F (raw bytes as read from flash):

Source 0x74–0x7F
199C 01 07 00 00 00 20 00 00 02 00 00 00
196B / 196C 02 00 00 00 00 40 00 00 01 00 00 00
After FlashTool's CF erase 00 00 00 00 00 40 00 00 01 00 00 00

A converted 196B only boots with the 199C values. With the 196 values it is completely dead.

Flash chips

Modules have either SST (0x00BF/0x2781) or ST M29W800 (0x0020/0x225B) chips. The programmer firmware uses SST's command sequences (unlock addresses 0x5555/0x2AAA). The M29W800 datasheet uses different unlock addresses, but the one ST module tried here accepted SST's sequences: it was erased, written and verified byte for byte five times with --force-st.

ST writes still need --force-st because that is one module in one programmer. The tool verifies every write, so a failed write is caught, but it leaves the module erased and in the programmer until you write it again from the backup. The programmer has no flow control, so the tools write one byte per serial call.

Cal bank (offsets within a bank)

Offset Meaning
0x20 Format header 0x0204
0x24 Length 0x1AB8
0x3E Checksum: u16 sum of bytes 0x44..0x1AB7
0x68 Model string (196B, 199C, …)
0x88 Serial number string
0x98, 0xA8 Free text (instrument ID, cal operator)
0xDA 0100 = 196, 0200 = 199 (probable)
0xDC Options: bit 0 Medical, bit 1 BusHealth
0x124, 0x154 0028 on a 196B, 0010 on color units
0x184 Language enable mask (one bit per language; not an option flag)
0x1AA 001E mono, 011E color
0x3FFC Write generation, counting down: the bank with the lower value is the live one

The meter rewrites the cal banks itself when it saves settings, and the padding between records picks up leftover RAM values. Two dumps of the same meter can therefore differ slightly in the cal area.

Options in firmware

  • Medical only changes the label on the version screen. With no option installed, one three-way setting can't take its third value; either option lifts that.
  • BusHealth is checked in about 12 places and enables the bus-health features.
  • Bonus Feature (5 extra screen memories) has no code check.

Firmware updates over the optical port

Fluke's FlashTool updates the firmware over the optical cable. It only offers COM1–COM8 and runs at 19200 baud by default (an update takes 10–20 minutes). Its data.bin is plain text: an init header, a RAM extension (32 × 512 bytes loaded at 0x44000000 with WB), then 484 blocks of the form \x1aDATA BLOCK\r\n,#H<addr>,#H<len>\r\n<len+1>\r\n<bytes><8-bit sum>\r\n. The blocks cover the bootloader and code, never the cal; the boot header (0x40000000, 0x28 bytes) is the last block.

The flow is roughly: ID, PC <baud>, EM FLUKEUHM (into the bootloader), upload the extension, save the cal to %temp%\<serial>.CAL, then the extension's commands (each a GO 0x44xxxxxx):

Command Does
CF Erases the whole chip, keeping only the cal banks, and writes a 16-byte struct at 0x70 from its own variables
PF <addr>,<len> Programs len bytes streamed from the port (no erase)
TI Checks the boot header: 0 valid, 1 not
EO Exit and reboot
FC, QF, FP Region report/query, port setup

Before trying it:

  • There is no sector erase, only CF. Any failure after CF leaves a meter that won't boot (the boot header comes last) and needs the programmer to recover.
  • On a converted 196B, CF writes 196-type values at 0x74–0x7C (table above), so the meter won't boot even if the update succeeds. They can't be fixed afterwards with PF, because changing 0 bits to 1 needs an erase.
  • Use the programmer to update a converted meter.

Battery and charger (for pack replacements)

Useful if you fit a Li-ion pack in place of the NiMH one:

  • The firmware charges with a NiMH profile: 1000 mA until a 6 h timer or dT/dt, 350 mA top-off for 2 h, then a 90 mA trickle. It has no voltage limit; the P-ASIC's VBATHIGH clamp sets one (47 µA into R4113, 38K3 for the stock ~11 V). A resistor in parallel with R4113 lowers it: roughly 0.285 V per kΩ of the combined resistance (120k in parallel → 29.0k → about 8.26 V measured).
  • A full charge only happens with the meter off. With the meter on, it trickles.
  • The gauge's capacity is 3500 mAh and only "learns" in 100 mAh steps, so a bigger pack reads empty long before it is. Charge is counted from the current setpoint, not the measured current, so a voltage-limited charge is overcounted.
  • Low-battery warning at 6.9 V, empty at 6.0 V.
  • The pack NTC is expected to be 10 kΩ at 20 °C. Charging starts only between 0 and 45 °C and stops at 50 °C. The pack's ID resistor (BATIDENT) sets the battery type.

Acknowledgements

The memory-module programmer (hardware, AVR firmware and the original programmer.py) is the work of the fluke19x-programmer project: github.com/feedbackloopchannel/fluke19x-programmer, with its video and the EEVblog Fluke 199C repair thread. convert_to_199c.py speaks that programmer's serial protocol; none of the module work here would have been possible without it. Thank you.

Hardware details (resistor values, charger behaviour, cal steps) come from Fluke's 192B/196B-C/199B-C service manual.

License

MIT, see LICENSE. Provided as is, with no warranty; see the caution at the top of this README. The programmer hardware and firmware are not part of this repository and are covered by their own project.

About

Tools for Fluke 19x ScopeMeters: options, 196->199C conversion, serial flash dump, FFT

Resources

Stars

0 stars

Watchers

0 watching

Forks

Releases

Packages

Contributors

Languages