This document describes the standard workflow for predicting chemical genotoxicity using the GenoToxMass platform, covering sample preparation, LC-MS/MS data acquisition, feature extraction, model invocation, result interpretation, and quality control.
It is intended for:
- Regulatory agencies
- Contract research organizations (CROs)
- Analytical laboratories
The procedure is designed to support standardized implementation, independent validation, and routine operational use.
This SOP defines the complete workflow for predicting chemical genotoxicity using the GenoToxMass platform, from sample preparation to result interpretation.
| Item | Specification |
|---|---|
| Recommended column | C18 column (2.1-3.5 um particle size, 2.1 x 50 mm) |
| Column temperature | 30 +/- 5 C |
| Mobile phase A | Water with 0.1% formic acid (LC-MS grade) |
| Mobile phase B | Acetonitrile with 0.1% formic acid (LC-MS grade) |
| Flow rate | 0.3-1.0 mL/min |
| Injection volume | 1-5 uL |
| Time (min) | %B |
|---|---|
| 0.0 | 5 |
| 1.0 | 5 |
| 8.0 | 95 |
| 10.0 | 95 |
| 10.1 | 5 |
| 12.0 | 5 |
| Item | Specification |
|---|---|
| Ionization source | Electrospray Ionization (ESI) |
| Scanning method | Select the mode exhibiting higher signal strength |
| Capillary voltage | 5.0 kV |
| Ion source temperature | 325 C |
| Desolvation gas flow | 12 L/min (N2) |
| Cone gas flow | 12 L/min (N2) |
| Collision energy | Ensure the molecular ion peak is retained at moderate intensity (10%-80%) |
| Mass range | m/z 50-1000 |
| Resolution | >= 20,000 (FWHM) at m/z 200 |
| Item | Specification |
|---|---|
| System suitability test mixture | Colchicine (CAS 64-86-8), Reserpine (CAS 50-55-5), each at 1 ug/mL |
| Acceptance criteria | Retention time variation <= 0.2 min; mass accuracy <= 5 ppm; intensity variation <= 20% across three consecutive injections |
- Filter sample through a 0.22 um nylon membrane filter.
- Add internal standard to a final concentration of 100 ng/mL.
- Concentrate 100 mL sample using solid-phase extraction (SPE).
- Condition SPE cartridge with 5 mL methanol, then 5 mL water.
- Load sample at 5 mL/min.
- Wash with 5 mL 5% methanol in water.
- Elute with 5 mL methanol.
- Evaporate to dryness under N2 at 40 C.
- Reconstitute in 200 uL methanol:water (1:1, v/v).
- Dissolve drug substance in an appropriate solvent at 10 mg/mL.
- Dilute to 10 ug/mL with methanol:water (1:1, v/v).
- Filter through a 0.22 um PTFE syringe filter.
- Homogenize sample (5 g) with 10 mL acetonitrile.
- Centrifuge at 10,000 x g for 10 min at 4 C.
- Collect supernatant and perform dispersive SPE cleanup.
- Evaporate 5 mL supernatant to dryness.
- Reconstitute in 500 uL methanol:water (1:1, v/v).
- Use data-dependent acquisition (DDA) or data-independent acquisition (DIA).
- For completely unknown samples, acquire both positive and negative ionization modes, then use the spectra from the mode with higher overall intensity for subsequent analysis.
- Acquire MS/MS spectra.
- Set dynamic exclusion to 5 s after two occurrences.
- Export raw data as
.raw. - Alternatively, use the GenoToxMass template (
.xlsx) with the following required columns:
| Column | Description | Example |
|---|---|---|
mz_list |
Comma-separated list of m/z values | 121.05, 149.02, 177.05 |
intensity_list |
Comma-separated list of corresponding intensities | 100, 45, 12 |
Software performs this section automatically.
- Identify the maximum intensity (
MaxM) in the raw spectrum. - Set filtering threshold
T = MaxM x P, whereP = 0.01(1%). - Retain only peaks with intensity >=
T.
| Feature | Abbreviation | Calculation / Definition |
|---|---|---|
| Peak Number | PN | Count of peaks after filtering |
| Base Peak | BP | m/z value of the most intense peak in filtered spectrum |
| Base Peak Proximity | BPP | m/z difference between the BP and its nearest peak |
| Maximum Mass | MaxM | Maximum m/z value in filtered spectrum |
| Maximum Mass Proximity | MaxMP | m/z difference between the MaxM and its nearest peak |
| Minimum Mass | MinM | Minimum m/z value in filtered spectrum |
| Mass Mean | MM | Mean of all m/z values in filtered spectrum |
| Mass Standard Deviation | MSD | Standard deviation of all m/z values |
| Intensity Mean | IM | Mean of all intensity values |
| Intensity Standard Deviation | ISD | Standard deviation of all intensity values |
| Intensity Density | ID | BPI / PN |
| Retention Time | RT | As acquired (minutes) |
| Collision Energy | CE | As acquired (eV) |
| Precursor Type | PT | 0 for [M-H]-, 1 for [M+H]+, [M+Na]+, [M+K]+ |
If the user does not provide the three test-related parameters (RT, CE, and PT), a generic input profile may be used:
RT = 8 minCE = 30 eVPT = positive ion mode
Important notes:
- If a specific ionization mode is selected, spectra must be acquired under that same mode.
- This fallback is not recommended unless absolutely necessary.
- Predictive performance under this mode is expected to degrade by >= 10% relative to external validation results.
- Navigate to
https://mspredict.com/genotoxicity. - Download the input template from the website.
- Prepare data following the template format.
- Upload the completed
.xlsxfile. - Click Predict and wait for processing (typically less than 5 seconds).
- Review the output:
- Prediction:
ToxicorNon-Toxic - Probability: value between
0and1
- Optimal threshold
tau*is determined by user need. - Classification rule:
- If probability >=
tau*-> classify as Toxic - If probability <
tau*-> classify as Non-Toxic
- If probability >=
| Application Priority | Recommended Threshold | Expected Performance Trade-off |
|---|---|---|
| Maximize Accuracy | 0.42 | Balanced sensitivity/specificity |
| Minimize False Negatives (Safety-First) | 0.30 | Higher recall, lower precision |
| Minimize False Positives (Cost-Sensitive) | 0.65 | Higher specificity, lower recall |
| Regulatory Confirmation | 0.55 | Optimized for both |
| Confidence Level | Probability Range |
|---|---|
| High | >= 0.8 or <= 0.2 |
| Medium | 0.6 <= p < 0.8 or 0.2 < p <= 0.4 |
| Low | 0.4 < p < 0.6 |
- Inject system suitability test mixture (colchicine, reserpine).
- Verify retention times are within +/- 0.2 min of established values.
- Verify mass accuracy <= 5 ppm for all three compounds.
- Run GenoToxMass prediction on all three.
- If any prediction deviates, investigate instrument performance.
- Include one positive control (for example, colchicine) and one negative control (for example, reserpine) in each analytical batch.
- Acceptance criteria: correct classification for both controls.
- If controls fail, reject the entire batch and re-analyze.
For laboratories implementing this method for the first time:
- Analyze the reference standard set (
n = 20 compounds). - Compare predictions with reference values.
- Acceptable performance:
>= 85%concordance. - Document all results for regulatory audit.
All analyses should be documented with:
- Instrument calibration records
- System suitability test results
- Sample preparation logs
- Raw data files
- Feature extraction worksheets
- Model prediction outputs
- Final interpretation and decision rationale
| Problem | Possible Cause | Solution |
|---|---|---|
| Low peak count after filtering | High noise level | Increase injection volume; improve sample cleanup |
| Unexpected predictions | Poor spectral quality | Verify collision energy; check precursor ion selection |
| Platform unavailable | Server maintenance | Contact support |
Converted from the uploaded SOP PDF for GenoToxMass.