Throughline

A traceability and reliability analyser for a medical device design history file.

It answers the question a design review actually turns on: can you draw an unbroken line from every user need, through the requirement that implements it, to the verification that tested it and the run that passed, and can you do the same for every hazard and its risk control. Where the line breaks, it says so and says how badly.

No dependencies outside the standard library. Python 3.9 or newer.

python3 fixtures/generate.py     # build the sample design file
python3 tests/run.py             # 90 tests
python3 bench/run.py             # writes results/benchmarks.json
python3 core/report.py --failures fixtures/failures.json --mission-hours 500

The report command exits non zero when there is a blocker, so it can sit in a pipeline and stop a release the same way a failing test does. An example report is in docs/example-report.md.

The shape of the thing

UserNeed  ->  Requirement  ->  Verification  ->  TestRun
                   ^
                Hazard  ->  RiskControl
Path What it does
core/model.py The objects, and the reverse index that makes the walk linear
core/load.py JSON in, objects out, with errors that name the offending entry
core/trace.py The twenty checks, and coverage and matrix generation
core/stats.py Incomplete beta, Clopper-Pearson, success run sizing, Weibull
core/report.py Markdown or JSON, plus the exit code

Two design decisions

Nothing raises during analysis. A design file is incomplete for most of its life, and a tool that refuses to load an incomplete file is useless exactly when you want it most. Loading rejects only what makes an object impossible to build, such as an unknown severity word. Everything else, including links that point at nothing, comes back as a finding with a severity attached.

Severity is graded, not flat. An unverified safety related requirement is a blocker; an unverified ordinary one is a major. A hazard with no control at all is a blocker; a control that exists but whose requirement was not flagged safety related is a minor. Sorting an undifferentiated findings list is how the important ones end up on page four.

Twenty rules in all. The ones that carry weight:

Rule Severity Catches
requirement-unverified blocker if safety related, else major a requirement nothing tests
verification-failed blocker the most recent run failed
contradictory-run blocker a run marked pass that reports failing units
impossible-run blocker more failures than units tested
hazard-uncontrolled blocker a hazard with no risk control
control-unverified blocker a risk control whose requirement is untested
risk-increased blocker residual risk worse than the initial risk
sample-size-short major a run with fewer units than the protocol requires
verification-not-run major a protocol that was never executed
dangling-*-link major a reference to something that does not exist
orphan-*, identifier-format minor, info hygiene

The check that matters most

latest_run takes the most recent run by date, not the best one. The fixture has a verification that went fail, then pass, then fail again after a firmware regression, and the report has to say fail. Taking the best result is the single most dangerous way to get this wrong, and it is a very easy thing to write by accident when you are reaching for any(r.result == "pass" ...). There is a test named after it.

Ties on the same date are broken by identifier so the answer is deterministic, which also has a test, because a report that changes between runs is a report nobody trusts.

Scoring the analyser instead of trusting it

The fixture generator plants eight specific defects and writes down what it broke, in fixtures/expected_findings.json. The analyser is scored against that list rather than against its own previous output:

  • every planted defect is found, with the expected severity
  • a clean design file produces zero blockers and zero majors
  • the number of serious findings does not exceed the number planted by more than two, because precision matters as much as recall and a checker that reports a hundred findings on a file with eight defects is not usable

A checker that reports nothing on a clean file has demonstrated nothing, so both directions are tested.

Statistics

Implemented from the definitions, with no third party library. The reason is not dependency minimalism: a reliability claim in a design history file has to be defensible line by line, and "the library said so" is not a defence. The tests check against published values and closed forms, not against a previous run.

  • Regularised incomplete beta, by the modified Lentz continued fraction. The binomial CDF and the Clopper-Pearson interval both go through it. Checked against direct binomial summation across 1,266 combinations: worst absolute error 2.6e-14.
  • Clopper-Pearson exact interval, one or two sided.
  • Success run sizing, which reproduces the two numbers every reliability plan quotes: 29 units for 90/95 and 59 for 95/95.
  • Weibull by median rank regression, with the r squared reported next to the parameters rather than buried, because a Weibull fitted to data that is not Weibull still returns two confident looking numbers.

The one sided versus two sided mistake

I wrote alpha/2 everywhere and the tests caught it, because the success run formula and the exact Clopper-Pearson bound stopped agreeing with each other.

They are two independent routes to the same number for a clean run, and when two independent routes disagree one of them is wrong. It was the interval. A two sided 95 percent interval puts 2.5 percent in each tail; a reliability demonstration quotes a one sided bound with the whole 5 percent in one tail. For a clean run of 29 units the upper bound on the failure rate is 0.0981 one sided against 0.1194 two sided. That is not rounding, it is the difference between demonstrating 90 percent reliability and demonstrating 88 percent.

clopper_pearson now takes a sided argument and reliability_lower_bound is explicitly one sided, with both closed forms pinned in the tests.

Scaling and accuracy

From results/benchmarks.json.

Analysis was quadratic and is now linear. Every reverse lookup was a list comprehension over the whole table, so asking "which verifications cover this requirement" once per requirement scanned everything twice over. It is invisible at fourteen requirements and very visible at eight hundred:

Requirements Objects Before After
50 191 1.2 ms 0.2 ms
100 386 2.8 ms 0.5 ms
200 782 2.9 ms 1.0 ms
400 1,555 10.7 ms 0.8 ms
800 3,084 42.6 ms 3.7 ms

Fitted scaling exponent went from 1.22 to 0.91, and the largest case got 11.5x faster. The fix is a lazily built reverse index that is discarded on any write, so it cannot drift from the tables it summarises.

How much data a reliability study actually needs. The Weibull recovery benchmark runs two ways on purpose. Fed the exact median rank quantiles it returns zero error to fifteen decimal places, which checks the algebra and proves nothing about behaviour on data. Fed genuinely random samples, 200 studies per cell:

True shape n = 10 n = 25 n = 50 n = 100
1.0 21.5% 13.2% 10.2% 8.4%
2.2 20.9% 13.9% 10.5% 8.3%
4.0 20.3% 13.7% 8.0% 6.8%

Median absolute error in the fitted shape parameter. The practical reading: a ten unit study estimates the Weibull shape to about 20 percent, so quoting a shape to two decimal places off ten units is false precision, and the interpretation that hangs off it, infant mortality against wear out, is not reliable at that sample size either.

Environment. Measured under CPython 3.12 compiled to WebAssembly, because no native interpreter was available on the machine I built this on. It is slower than native, and the runtime clamps perf_counter to a 0.1 ms floor, measured in the benchmark rather than assumed. Several per operation timings sit at that floor and are the clock rather than the code; they are left in the JSON and not quoted here. The scaling table and the accuracy figures are unaffected, because both are comparisons rather than absolute times.

How it is tested

90 tests, about 0.15 seconds.

tests/test_trace.py   36   the rules, the matrix, the planted defect scoring
tests/test_stats.py   33   incomplete beta, exact intervals, success runs, Weibull
tests/test_load.py    21   model validation, loader errors, multi file merge

Known gaps

  • Synthetic data throughout. The device, the requirements and the failure times are invented, chosen so the tests can assert against known answers.
  • Risk is modelled as an ordered severity and probability pair. There is no risk acceptability matrix, because that is a per organisation policy document rather than something a tool should assume.
  • No coverage of design validation, only verification. Validation asks whether the right device was built, and that is not a question you answer from a link graph.
  • Median rank regression rather than maximum likelihood for the Weibull fit. It is what reliability standards describe and it yields the r squared, but MLE has better small sample properties.
  • Single file or single directory input. No database, no concurrent editing.