Controller information and settings

All methods use the existing native TCP session and share the operation lock with measurement control. Nothing in this section is applied automatically on startup or reconnection.

Method Meaning and limits
controller_info() Read the controller GUID, serial number, production time, firmware version label and raw RTC status. Nonzero rtc_status_raw requests clock setting; individual bits are unknown.
unit_info(unit) Read a discovered unit's GUID, serial number, production, validity and latest calibration times and firmware label, unit 1..32.
history_minimum_index() Read the oldest history position the controller can still resend. Read-only.
set_upload_interval(interval_raw) Change periodic status uploads for the whole controller, in seconds (1..255). Zero is unsupported. This does not change recipe recording rules or history.
set_clock(local_time) Set an explicit naive datetime in controller-local time, years 2000..2099 with millisecond precision. No timezone inference or conversion.
initialize_test_id(seed, source=...) Advanced device-wide maximum TestID synchronization, unsigned 32-bit value. Requires a nonempty description of the complete inventory or database establishing that maximum.

ControllerInfo and UnitInfo are frozen dataclasses. They decode the serial-number replies (002A/002B, see commands) and keep the complete replies and firmware-version payloads as bytes. Our controller reports factory placeholder serial numbers. A firmware version query only reads information; firmware modification is not exposed.

Clock and TestID changes require complete discovery and status within the configured number of upload intervals for every discovered channel, all finished and alarm-free. Paused, stopped, running, missing, stale and unknown channel states do not satisfy this requirement. These checks include channels without an application profile. TestID seeds below any locally known run or status TestID are rejected; local observations alone cannot prove the device-wide maximum.

Successful setting calls mean the device acknowledged the write. TestID initialization additionally requires the returned value to equal the seed. Clock readback, setting persistence and counter wrap behavior remain unknown. Lost or malformed acknowledgements close the connection and raise OperationUncertain; there is no automatic retry. Explicit rejections raise OperationRejected. Reads and writes are serialized without stopping status or measurement reception.

The interval is in seconds. Freshness is upload_interval_raw (default 5) multiplied by missed_upload_intervals (default 3), configured on Client. The initial interval must match the controller's setting; constructing a client does not write device settings. An acknowledged interval change updates the freshness threshold immediately. Supply the new interval on subsequent restarts.

from datetime import datetime

info = await client.controller_info()
unit = await client.unit_info(1)
await client.set_upload_interval(2)
await client.set_clock(datetime(2026, 9, 18, 14, 30))

Choose the clock value explicitly for the installation. These examples do not initialize TestIDs; that requires a separately established complete inventory.