Cary 630 FTIR USB protocol

A best-effort description of the USB protocol of the Agilent Cary 630, as implemented by the driver in caryftir/. It is based on Wireshark captures of MicroLab PC. Each part is marked:

  • observed: seen in the captures;
  • expected: part of the protocol, but not exercised in the captures;
  • guess: a meaning inferred from values alone.

To decode a capture:

python3 tools/decode_frames.py capture.pcapng 60,61,62,63,64   # commands + replies
python3 tools/decode_tlv.py    capture.pcapng                  # value lists on EP 0x85

Old driver hang

The instrument takes one scan and then waits until the host has fetched and released it (observed: every scan in every capture, without exception):

0x60 POLL_SCAN        repeat until status '@' (scan ready)
0x62 SELECT_BLOCK 0x4000 → 0x64 READ_BLOCK → 0x63 RELEASE_BLOCK     (scan info)
0x62 SELECT_BLOCK 0x0001 → 0x64 READ_BLOCK → 0x63 RELEASE_BLOCK     (raw interferogram)
0x61 RELEASE_SCAN

The old driver left out the second 0x63 and the 0x61, so the instrument never started the next scan. It also polled 0x60 a fixed number of times instead of waiting for '@'.

Endpoints (observed)

EP Dir Use
0x04 OUT Commands, always 64-byte packets
0x83 IN Replies, always 64-byte packets
0x06 OUT Messages longer than 60 bytes, announced by frame type 0x0A
0x85 IN Data (value lists, blocks), announced by frame type 0x0B

Framing

Every packet on EP 0x04 and 0x83 has the following layout:

[type][seq][len_hi][len_lo] [message, len bytes] [zero padding to 64]
Type Meaning
0x08 command, message ≤ 60 bytes observed
0x09 reply observed
0x0A host announces a longer message: body = BE32 length. The device answers with an empty 0x0A, the message follows on EP 0x06 in ≤512-byte pieces, then a normal 0x09 reply arrives observed
0x0B device announces data on EP 0x85: body = BE32 total length observed
0x0C device rejects the frame expected
0x0D link reset, the first frame of a session. Reply body: 4 bytes (07 ff 00 01) observed

The sequence byte increments once per message and is echoed in the reply.

Messages

Every command message starts with [command][address][0][n], followed by fields. All multi-byte fields are big-endian. The address byte is 0 except where noted. n is non-zero only for 0x15, where it is 8.

In replies, byte 2 is the status:

Status Meaning
'@' 0x40 OK
'A' 0x41 error
'C' 0x43 still scanning
'D' 0x44 last chunk of a block
'E' 0x45 failed
'`' 0x60 busy

The original software leaves leftovers from earlier messages in unused message bytes. Those bytes carry no meaning, and much of the "magic" hex in the old driver came from them.

Every message on EP 0x85 starts with an 8-byte header: [command][index][status][0x10][BE32 payload length].

Commands

Cmd Name Request Reply
0x01 CONNECT bytes 4–5 = 01 2a '@' observed
0x19 CONNECT_CONFIRM – '@' observed
0x68 GET_DATA_TYPES – bytes 4–7 = data types available per scan, 8–11 = available as averages (this instrument: 0x4111, 0x0110) observed
0x18 READ_VALUE_LIST address = scan index, bytes 4–7 = list (0x4000 scan info, 0x2000 system info) 0x0B + value list on EP 0x85 observed
0x69 REQUEST_DATA_TYPES bytes 4–7 = data types wanted per scan, 8–11 = averaged types wanted '@' observed
0xB2 GET_VALUE bytes 4–7 = tag (48-byte message) bytes 8–9 = value length, value from byte 16 observed
0xB3 SET_VALUE bytes 4–7 = tag, 8–9 = length, 12–15 = 0 or 1 (meaning unknown), value from byte 16 '@' observed
0xB4 SET_VALUES bytes 4–5 = size of the value list, 9 = transaction id, value list from byte 16 '@' observed
0xB6 REFRESH_INFO byte 9 = transaction id '@' observed
0xB5 POLL_TRANSACTION byte 9 = transaction id '@' done, '`' busy observed
0x15 CONTROL byte 4 = control id, 5 = 1, 8–9 = data length, 10–11 = 4, data from byte 16 '@' observed
0x60 POLL_SCAN – 'C', or '@' with bytes 4–7 = scans completed, 8–11 = data types in this scan, 12–15 = averaged types, 16–19 = run id observed
0x62 SELECT_BLOCK bytes 4–7 = data type, byte 11 = 1 for the averaged version bytes 4–7 = element count, 8–11 = element size observed
0x64 READ_BLOCK address = scan index 0x0B + one chunk on EP 0x85. Repeat until a chunk has status 'D' observed for single chunks; multi-chunk expected
0x63 RELEASE_BLOCK – '@' observed
0x61 RELEASE_SCAN – '@' observed
0x02 DISCONNECT – '@' observed
  • Transaction id: one counter shared by 0xB4 and 0xB6. It starts at 1 and increments per message.
  • Scan index: starts at 0 after connecting and increments once per scan. The host keeps it, and it matches "scans completed − 1".
  • Other commands: 0xB0 and 0xB7 appear in the captures (around gain changes); 0xA4, 0xB8 and 0xB9 are expected. The driver does not use them.

Blocked: 0xB1 WRITE_NONVOLATILE

0xB1 writes values the instrument keeps across power cycles, such as calibration data. It is known and deliberately blocked: protocol.BLOCKED_COMMANDS lists it, and the transport layer raises BlockedCommandError.

For safety reasons the implementation of this has intentionally been deferred. We recommend using the original software for any non-volatile writes and adjustments.

Control ids (0x15)

Id Name
0x10 START_SCANNING observed, sent after configuring
0x11 STOP_SCANNING observed, sent before disconnecting
0x20 START_BACKGROUND expected
0x21 START_SAMPLE expected

The observed stop message also has non-zero bytes 7 and 12–15. These differ from the start message and appear to be don't-care.

Data types

Data-type ids are used in 0x62 and as bits in the masks of 0x68, 0x69 and 0x60.

Bit Name Elements
0x0001 RAW 4 bytes: class byte + 24-bit signed BE value. Even class = IR detector, odd class = reference laser. Trailing ffffffff records are unused observed
0x4000 SCAN_INFO value list describing the scan observed
0x0002 / 0x0004 RAW_IR / RAW_LASER BE int32 expected
0x0010 INTERFEROGRAM (resampled) BE int32 expected
0x0040 VELOCITY BE float32 expected
0x0100 SINGLE_BEAM BE float32. Wavenumber range from tags 0x3904–0x3914 (guess) expected
0x0200 MINI_SINGLE_BEAM BE float32 expected
0x0400 SPECTRUM BE float32 expected

The original software only requests RAW (0x69 with masks 1, 0) and computes spectra on the PC. However, the instrument reports that it can deliver INTERFEROGRAM and SINGLE_BEAM itself, both per scan and averaged (0x68 → 0x4111 / 0x0110). The host also sends spectrum settings (apodization, phase correction). Requesting SINGLE_BEAM may therefore give finished spectra without host-side processing. This is untested.

Values (tags)

Instrument values are addressed by 16-bit tags. The low nibble appears to encode the type: 4 float32, e string, otherwise integer (guess, consistent with every observed value). In value lists each value takes 6 bytes: BE16 tag + BE32 value.

Tag Name Observed values
0x0031 RUN_ID: incremented each time a run starts; scans report it 1, 2, 3 … observed
0x0041 RUN_MODE 1 observed
0x0101 RESOLUTION (cm⁻¹) 4 observed
0x0111 PHASE_POINTS 512 observed
0x0121 / 0x0131 LEFT_PAD / RIGHT_PAD 128 / 200 observed
0x0141 AVERAGE_COUNT: scans co-added 1, 3, 4, 8 observed
0x3101 APODIZATION 1, 5 observed
0x3111 ZERO_FILL 0 observed
0x3121 OFFSET_CORRECTION 0 observed
0x3131 PHASE_CORRECTION 1 observed
0x3141 PHASE_APODIZATION 1 observed
0x1041 SCAN_MODE: 4 = normal 4 observed
0x0211 unknown. Not the scan rate: the time between scans does not follow it. Not used by the driver 25, 50 observed
0x0231 INTERFEROGRAM_PEAK; reset to 0 before starting e.g. 0x3d5cd8 observed
0x0221, 0x0241 unknown; reset to 0 before starting 0 observed
0x102e SERIAL_NUMBER MY2446CU14 observed
0x130e–0x136e version / part strings observed
0x1201 / 0x1211 IR_GAIN / LASER_GAIN 212 / 204 observed
0x3154 MAX_WAVENUMBER 7630.0 observed
0x3184 LASER_WAVENUMBER: reference-laser wavenumber at 0 °C (calibration offset) 7659.79 educated guess
0x3194 LASER_TEMPERATURE_COEFFICIENT: cm⁻¹ per °C (calibration slope) −0.47 educated guess
0x31a1 AXIS_CORRECTION_OFF: 1 disables the temperature correction in the original software 0 educated guess
0x31c1 AXIS_CORRECTION_APPLIED: which correction the original software applied 2 educated guess
0x1901 SHUTDOWN_FLAGS: bit 0x01 instrument shutting down, bit 0x10 battery critical 0 expected
0x1911 SAMPLING_ACCESSORY: changes when the sampling accessory changes 0 expected
0x1944 CPU_TEMPERATURE (°C) 52.0 expected
0x1954 POWER_TEMPERATURE (°C) 51.9 expected
0x1964 IR_TEMPERATURE (°C) 59.6 expected
0x1974 DETECTOR_TEMPERATURE (°C) 40.0 expected
0x1984 BLOCK_TEMPERATURE (°C); also used for the laser temperature correction 49.25 expected
0x1934, 0x1994, 0x19a4, 0x19b4 further temperature slots; 999 = no sensor 999 expected
0x19e4 / 0x19f4 SOURCE_VOLTAGE / SOURCE_CURRENT (IR source) 2.70 / 1.86 expected
0xd814 unknown float; rose from 40.7 to 54.1 within one 75 s capture observed
0x3904 / 0x3914 FIRST/LAST_WAVENUMBER of instrument-computed spectra guess

The resolution sets the interferogram length: at 4 cm⁻¹ a RAW block holds 58 515 records.

Laser temperature correction (educated guess)

The wavenumber axis depends on the reference laser's wavenumber, which drifts with temperature. The best reading of the values above is:

laser wavenumber = 0x3184 + 0x3194 × 0x1984      (7659.79 − 0.47 × 49.25 ≈ 7636.6 cm⁻¹)

The original software appears to compute the laser wavenumber like this from the block temperature, then put spectra on a fixed grid of k · 2 · 7633.588 / N cm⁻¹. 0x31a1 and 0x31c1 only control and record this correction on the PC. They don't change the measurement, so the driver doesn't write them.

The captured data can't confirm the correction. At 4 cm⁻¹ the apparent peak of the CO₂ Q-branch isn't known to better than a few tenths of a cm⁻¹. In captured backgrounds it lands at 667.79 cm⁻¹ with the correction and 667.52 without it (band centre 667.38). An earlier version used 0x1964 as the temperature, but that is the IR temperature, not the block.

caryftir.processing applies the correction by default (ProcessingSettings.temperature_correction). It falls back to 7633.588 cm⁻¹ if the values are missing or implausible.

Status, warnings and errors

The instrument doesn't send error codes for its condition. There are two mechanisms, both visible in every scan:

  1. Shutdown (instrument side). 0x1901 bit 0x01 means the instrument is shutting down, and bit 0x10 means a critical battery. The original software closes the connection. The driver raises InstrumentShutdown from next_scan(), after the scan has been released.
  2. Limits (host side). Everything else, including the original software's "overheating" message, is the PC comparing readings against limits. The default limits come from the original software's configuration file (MicroLabPC.exe.config in its installation folder).

Results are a warning (the original's "marginal") or an error ("critical"). The original software shows them as a status and keeps going.

Check Reading Warning outside Error outside
block temperature 0x1984 −10 … 70 °C −20 … 80 °C
detector temperature 0x1974 38 … 42 °C 36 … 44 °C
energy peak |IR| of the raw scan ÷ 256 22 000 … 28 000 20 000 … 30 000
laser peak |laser| of the raw scan ÷ 256 5 500 … 15 000 3 000 … 17 000
sampling accessory 0x1911 any change
  • Switched off by default: the configuration also holds limits for the CPU (43 … 44 / 42 … 45 °C), IR and power temperatures (18 … 47 / 15 … 50 °C), but the original software has them switched off. Captured IR is 59.6 °C.
  • Not configured: the original software also has fields for source voltage and current limits, but none are set in the configuration.
  • Energy: the original software aims for an energy of 26 000 by adjusting the IR gain (0x1201). The driver doesn't adjust the gain. With the gain in the captures, energy reads about 15 800, so every scan gets an energy error with the default limits. Lower the limit or adjust the gain if that's expected.
  • Confidence: the meanings of the status tags come from how the original software labels and checks them (marked "expected" in the tag table). The readings and their values are observed.

A session

  1. Connect: 0x0D reset, 0x01, 0x19, 0x68, 0x18 scan info, 0x18 system info.

  2. Start:

    1. 0xB3 SCAN_MODE = 4
    2. settings (0xB4 with the ten values above, 76 bytes, so it goes via EP 0x06)
    3. 0xB3 peak and unknown values = 0
    4. 0x69 (requested data types)
    5. 0xB4 new run id
    6. 0x15 START_SCANNING
    7. 0xB6, then 0xB5 until done

    The original software additionally reads many values and writes 0x0211 (purpose unknown), 0x31a1 and 0x31c1 (temperature correction flags, see above). The driver skips these; none of them changes the measured data.

  3. Scan loop: see "Why the old driver hung".

  4. New settings mid-session: send the settings again plus a new run id. Scans then report the new run id.

  5. Stop: 0x15 STOP_SCANNING, then 0x02.

Open questions

  • What do the other class-byte bits in RAW records mean? Values 0, 1, 4, 8, 9, 16, 17 … occur.
  • What do the 0xB3 flag (bytes 12–15) and tags 0x0211, 0x0221 and 0x0241 do?
  • Does the laser temperature correction hold?
  • The energy limits: with the gain used in the captures, energy is about 15 800, below the default error limit of 20 000 (see "Status, warnings and errors").
  • Lots of work to confirm we get the same data as the original software.